| INPUT
RESISTANCE: |
100 GigaOhms |
|
MAXIMUM ELECTRODE IMPEDANCE: |
300 MegaOhms |
|
EQUIVALENT INPUT NOISE (wideband): |
X1000 Gain <5 microvolts rms; X5000 Gain
<2.5 microvolts rms; X10000 Gain <2.5 microvolts rms |
| CMMR: |
60 Hz >90dB; 120Hz >85dB |
| SIGNAL
TO NOISE RATIO: |
>110dB |
| GAIN: |
continuously variable over three ranges
(+/-5%) : 0-1000, 0-5000, 1-10000 |
| D.C.
OUTPUT OFFSET: |
continuously variable (switched in) : +-14
Volts |
| LOW FREQUENCY
CUTOFF IN A.C. MODE: |
0.5 Hertz |
| FREQUENCY RESPONSE
(+/- 10 Volts out): |
D.C. to 10 kHz +/-5% |
| OUTPUT HIGH
FREQUENCY CUTOFFS (Hz +/-5%): |
100, 250, 500, 1000, 2500,
5000, OUT |
| OUTPUT
LOW FREQUENCY CUTOFFS (Hz +/-5%): |
OUT, 0.5, 2, 20, 100, 200 |
|
ATTENUATION BEYOND CUTOFFS: |
12 dB per octave; 40 dB per decade Active
Butterworth Filters |
| OUTPUT VOLTAGE
(maximum): |
+- 14 Volts/channel |
| OUTPUT IMPEDANCE: |
Approximately 50 ohms |
| OUTPUT RESISTANCE: |
50 Ohms/channel |
| POWER SUPPLY: |
85-265 VAC 50-60 Hz, auto-switching |
| FUSE: |
1 Amp quick blow |
|
ENCLOSURE: |
Flame retardant ABS, UL flammability rating
94V-O |
|
DIMENSIONS: |
11" x 7.75" x 3" (280mm x 200mm x 75mm) |
| WEIGHT: |
Approx. 3lbs (1.4 kg) |
| WARRANTY: |
5 years |
| NOTE: |
This device is not isolated. All
non-isolated amplifiers are not intended for use with human
subjects. For human subjects, please select an ISOLATED AMPLIFIER such as
Model
2024F. |