Intronix Model 2024F
Electrophysiological Amplifier System with Signal Conditioning

$5100.00 U.S. Dollars

Specifications

 
 
INPUT RESISTANCE: 100 GigaOhms
MAXIMUM ELECTRODE IMPEDANCE: 300 MegaOhms
EQUIVALENT INPUT NOISE (wideband): X1000 Gain <5 microvolts rms;  X5000 Gain <2.5 microvolts rms; X10000 Gain <2.5 microvolts rms
CMMR: 60 Hz >90dB; 120Hz >85dB
SIGNAL TO NOISE RATIO:  >110dB
GAIN: continuously variable over three ranges (+/-5%) : 0-1000, 0-5000, 0-10000
D.C. OUTPUT OFFSET: continuously variable (switched in) : +-14 Volts
LOW FREQUENCY CUTOFF IN A.C. MODE:  0.25 Hertz
FREQUENCY RESPONSE (+- 5 Volts out): D.C. to 10 kHz (+/-5%)
OUTPUT HIGH FREQUENCY CUTOFFS (Hz +/-5%): 100, 250, 500, 1000, 2500, 5000, OUT
OUTPUT LOW FREQUENCY CUTOFFS (Hz +/-5%): OUT, 0.5, 2, 20, 100, 200
ATTENUATION BEYOND CUTOFFS:  12 dB per octave; 40 dB per decade Active Butterworth Filters
OUTPUT VOLTAGE (maximum): +- 14 Volts/channel
OUTPUT IMPEDANCE: Approximately 50 ohms
OUTPUT RESISTANCE: 50 Ohms/channel
ISOLATION, input/output: 2500 VRMS 60Hz
LEAKAGE CURRENT, input/output: <2uA at 120 VAC 60Hz
POWER SUPPLY: 85-265 VAC 50-60 Hz, auto-switching
FUSE: 1/2 Amp Slo-Blow
ENCLOSURE: Flame retardant ABS, UL flammability rating 94V-O
DIMENSIONS: 11" x 7.75" x 3" (280mm x 200mm x 75mm)
WEIGHT: Approx. 3lbs (1.4kg)
WARRANTY: 5 years
 

All specifications are subject to change without notice

 
Copyright 2008 [Intronix Technologies Corporation]. All rights reserved